Submit the complete original model and installation requirements for an independent mechanical, electrical and calibration review. Source facts below identify the original instrument; proposed PPS compatibility is confirmed separately.

Specify this configuration ↘

Technical content reviewed 16 November 2026 · PiezoPressureSensor

PCB Piezotronics 113B, 102B, 137B & 138A Replacement Review: sensing and signal-path concept
Original engineering concept; configuration-specific product drawings are reviewed separately.

01 / ENGINEERING NOTES

Distinguish the pressure mount from the electrical connector

For the exact 102B reference, 3/8-24 is the mounting thread and 10-32 is the coaxial electrical jack. The 113B22 reference uses a separate floating clamp arrangement. A shared electrical connector does not establish the same pressure-port geometry.

02 / ENGINEERING NOTES

Incident and reflected rise times are different fields

The 137B23B original pencil probe specifies incident rise time ≤6.5 μs. The 113B22 and 102B references describe a different response/geometry context. Do not apply one family’s microsecond number to every original-model alternative.

03 / ENGINEERING NOTES

Keep the underwater medium with the exact suffix

The 138A05 reference uses tourmaline in an underwater ICP assembly. Hydrostatic baseline, transient range, cable sealing and temperature require independent checks. Piezoresistive probe photographs are not evidence of a quartz or tourmaline equivalent.

04 / ENGINEERING NOTES

Four acceptance dimensions before a written quotation

Use the source table below as a reference input. Send your full original suffix and port drawing, then define the acceptance limits.

  • Pressure: reference, baseline, dynamic range and proof/survival limits.
  • Mechanical: thread pitch, diaphragm location, seal shoulder/O-ring and cable routing.
  • Thermal/media: operating versus compensated temperature, flash exposure and wetted compatibility.
  • Electrical: output type, excitation, sensitivity, conductor/pinout and usable installed response.

ORIGINAL-MODEL DATA / INDEPENDENT REVIEW

Keep original specifications separate from a proposed replacement.

Facts below describe the named original models. A PPS configuration still needs its own dimensional, electrical and calibration evidence. Resonance and flat bandwidth are different quantities.

Original reference facts checked 16 November 2026
Original model & sourceVerified reference factsCompatibility checkReview request
PCB Piezotronics 113B225,000 psi / 344.7 bar; 1 mV/psi; resonance ≥500 kHz; rise time ≤1 μs.The range is for ±5 V output. Resonance is not flat bandwidth. Mounting uses a separate floating clamp nut; electrical 10-32 is not a pressure-port thread.Review this original ↘
PCB Piezotronics 102B5,000 psi / 344.7 bar; 1 mV/psi; 3/8-24 mount; 10-32 electrical connector.Do not conflate mechanical 3/8-24 mounting with electrical 10-32 connector. M/S suffix variants require their own verification.Review this original ↘
PCB Piezotronics 137B23B50 psi / 3.447 bar; 100 mV/psi; incident rise time ≤6.5 μs; BNC.Incident rise time is 6.5 µs, not the ≤1 µs reflected response of 113B/102B. No general sub-microsecond claim is justified across blast families.Review this original ↘
PCB Piezotronics 138A055,000 psi / 344.7 bar; tourmaline ICP; reflected rise time ≤1.5 μs.Water coupling, hydrostatic loading and exact suffix matter; this is not evidence for a PPS underwater rating.Review this original ↘

No affiliation or universal interchangeability is implied. Pressure-port threads and electrical connector threads must be checked separately.

Pressure-chain questions for this configuration

Is this a guaranteed drop-in equivalent?

Compatibility is confirmed only after the complete original variant, port drawing, electrical interface and proposed-unit evidence have been reviewed.

What should I attach first?

The original datasheet, port or installation drawing, and required range/response/temperature are the most useful starting documents.

Sources and scope of the comparison

Original-model facts and general methods describe their named sources. They do not qualify a PPS unit or imply manufacturer affiliation.

COMPLETE THE PRESSURE MEASUREMENT CHAIN

Specify conditioning, mounting and protection together.

Optional budget requests. Confirm the input type, response, channel count and installed compatibility. Specialist hydrophone preamplifiers, hot-zone hardline cables and cooled assemblies are quoted separately.

Review selected configuration ↘

PREPARE A REVIEWABLE BRIEF

Download the worksheets for your project.

These are blank engineering request/checklist documents. They are not released product datasheets, calibration certificates or signed quotations.

ENGINEERING RFQ / PCB PIEZOTRONICS / ORIGINAL-MODEL REVIEW

Send this pressure configuration for review.

Include the original suffix and a private installation drawing when available. Final specifications, evidence, price and lead time are confirmed in writing.

PPS replacement review

Pressure event & project contact
Document package requested

Optional requests for the exact offered assembly. Availability, scope and approval are confirmed during quotation.

Accessory requests for this chain

Opt in to the items you need. These are budget requests; compatibility, channel count and final price require review.

Bring the whole pressure event into focus.

Discuss your application ↗