01 / ENGINEERING NOTES

M5×0.8 and M5×0.5 are different interfaces

The current XTL-190(S/L)(M) PDF identifies 10-32 UNF imperial versions and M5×0.8 metric versions. Specify the actual suffix and port drawing. A replacement must match thread pitch, engagement and sealing location separately.

02 / ENGINEERING NOTES

Do not infer liquid compatibility from a metal package

The XTL-190 source explicitly says performance is not guaranteed in water and requests factory review of medium and O-ring compatibility. Record the actual gas/liquid and isolation need before considering an equivalent.

03 / ENGINEERING NOTES

Thin and water-cooled packages change the review

The LQ-062 original is a nominal 0.76 mm thin-line package. WCTV references are cooled assemblies with specific output, coolant and installation conditions. Exact WCT/WCTV/EWCTV identifiers are not interchangeable.

04 / ENGINEERING NOTES

A Wheatstone bridge needs its own electronics

Nominal bridge output, rated excitation, zero balance and compensated versus operating temperature are separate specifications. Confirm cable conductors, gain and both pressure-path responses when using a differential reference.

05 / ENGINEERING NOTES

Four acceptance dimensions before a written quotation

Use the source table below as a reference input. Send your full original suffix and port drawing, then define the acceptance limits.

  • Pressure: reference, baseline, dynamic range and proof/survival limits.
  • Mechanical: thread pitch, diaphragm location, seal shoulder/O-ring and cable routing.
  • Thermal/media: operating versus compensated temperature, flash exposure and wetted compatibility.
  • Electrical: output type, excitation, sensitivity, conductor/pinout and usable installed response.

ORIGINAL-MODEL DATA / INDEPENDENT REVIEW

Keep original specifications separate from a proposed replacement.

Facts below describe the named original models. A PPS configuration still needs its own dimensional, electrical and calibration evidence. Resonance and flat bandwidth are different quantities.

Original reference facts checked 16 November 2026
Original model & sourceVerified reference factsCompatibility checkReview request
Kulite XCQ-062Four-arm bridge; 10 V excitation, nominal 100 mV full scale; range-dependent natural frequency.Natural frequency without screen is range-dependent (150–1000 kHz); it is not a universal flat bandwidth.Review this original ↘
Kulite XTL-190(S/L)(M)10 V bridge; 10-32 UNF or M5×0.8 variants; −55 to +175 °C / −67 to +347 °F.PDF explicitly says performance is not guaranteed in water; consult factory for medium/O-ring compatibility. M5×0.8 and M5×0.5 must be separate configurations.Review this original ↘
Kulite LQ-0620.030 in / 0.76 mm nominal thin package; bridge; LQ temperature −55 to +120 °C.LQ and LE temperature ratings differ; do not apply the LE limit to LQ. Screenless natural frequency is range-dependent; no flat DC–150 kHz guarantee.Review this original ↘
Kulite WCTV-312(M)Cooled silicon bridge; nominal 100 mV full scale. Cooling conditions and exact suffix required.Use exact WCTV versus WCT/EWCTV naming. Gas/media temperature, internal sensor temperature and cooling conditions are different. Do not copy 1093°C as the silicon-chip operating temperature.Review this original ↘

No affiliation or universal interchangeability is implied. Pressure-port threads and electrical connector threads must be checked separately.

Pressure-chain questions for this configuration

Is this a guaranteed drop-in equivalent?

Compatibility is confirmed only after the complete original variant, port drawing, electrical interface and proposed-unit evidence have been reviewed.

What should I attach first?

The original datasheet, port or installation drawing, and required range/response/temperature are the most useful starting documents.

Sources and scope of the comparison

Original-model facts and general methods describe their named sources. They do not qualify a PPS unit or imply manufacturer affiliation.

COMPLETE THE PRESSURE MEASUREMENT CHAIN

Specify conditioning, mounting and protection together.

Optional budget requests. Confirm the input type, response, channel count and installed compatibility. Specialist hydrophone preamplifiers, hot-zone hardline cables and cooled assemblies are quoted separately.

Review selected configuration ↘

PREPARE A REVIEWABLE BRIEF

Download the worksheets for your project.

These are blank engineering request/checklist documents. They are not released product datasheets, calibration certificates or signed quotations.

ENGINEERING RFQ / KULITE / BRIDGE AND PACKAGE REVIEW

Send this pressure configuration for review.

Include the original suffix and a private installation drawing when available. Final specifications, evidence, price and lead time are confirmed in writing.

PPS replacement review

Pressure event & project contact
Document package requested

Optional requests for the exact offered assembly. Availability, scope and approval are confirmed during quotation.

Accessory requests for this chain

Opt in to the items you need. These are budget requests; compatibility, channel count and final price require review.

Bring the whole pressure event into focus.

Discuss your application ↗